The model 828 AC /D C V-A Source (figure 1) is a
compact, seff-regulating voltage and Current source,
capable of meeting the calibration needs of most
measuring instruments and suitable for use in the
laboratory, on the production line, or in the field.
Three sets of interlocking pushbutton switches
permit output mode and range selection. Mode selection
provides either dc output or one of three internally
generated ac frequencies. If desired, an external 50-440 Hz
signal source may also be used to
determine the output frequency. There are four voltage
ranges and five current ranges, and the output
can be set to a percentage of the selected range value
by a set of thumbwheel switches.
The marked value of all ranges can be exceeded by
100%, except the 1000V range which is limited to
1100V maximum, and the 10A range, which is
limited to 5 amperes maximum {7 amperes on special
order).
Two pushbutton switches on the front panel apply
and remove the output signal from the output terminals.
Pressing the OPERATE switch applies the output signal to
the terminals and lights a green indicator lamp; pressing
the DE-ENERGIZE switch removes
the signal and lights a red indicator lamp. The red
lamp will also light whenever the internal protection
circuitry automatically removes the output signal
from the front panel terminals.
A three-position toggle switch allows selection
between normal, run up, or percent deviation operation of
the model 828. In the NORMAL position, the
output present at the output terminals is exactly the
value set by the thumbwheel switches and the range
selector, in the RUN UP position, a rotary control
permits the operator to vary the output from zero to
102% of the value set by the thumbwheel switches;
this allows analog meters to be examined for frictional
problems in their movements. In the % DEV
position, a calibrated control varies the output
between 98 and 102% of the value set by the thumb
wheel switches and provides a direct reading of the
percent error of the device under test.
Hасто́льная кни́га тип:
Обслуживание и Руководство пользователя
Страницы:
28
Размер:
1.53 Mbytes (1605325 Bytes)
Язык:
english
Пересмотр:
Hасто́льная кни́га ID:
Дата:
Качество:
Сканированный документ, читая частично плохо, частично не читается.
Дата загрузки:
2017 04 16
MD5:
24f5b7686a29fe9f007f63279d44ec68
Загрузки:
679
Информация
Title Page
...
... i
W
arranty ... ii
High Voltage Warning and Safety S u m m a
ry ...
... iii
Table of Contents
...
... v
List of Illu s tra tio n s
...
... vi
List of
Tables ...
... vt
D E S C R IP T IO N
... 1
SPECIFICATIONS
...
... 2
PRINCIPLES OF O P E R A TIO N
...
... 2
OPERATION INSTRUCTIONS
...
... 4
Controls and Indicators
...
... 4
Initial Set u p
...
6
G
eneral ...
...
6
Voltage S
election ...
6
External C
onnections ...
... 7
O
peration ...
7
W a rm -U p
...
... 7
Basic Operating Procedure
...
... 8
Errors at Low Output
Voltage ...
8
CALIBRATION A N D M A IN T E N A N C E
...
C
alibration ...
Equipment
Required ...
S e tu p
...
...
Calibration
Procedure ...
Maintenance and R
epair ...
THEORY OF O P E R A T IO N
...
Output Signal
Level ...
Constant-Current S o u rc e
...
In te grato
r ...
Output A m p lifie
r ...
Overload
Protection ...
Output Transform
er ...
S w itch in g
...
Run Up and % Deviation
...
Turn-On D e la
y ...
Power S u p p
ly ...
CALIBRATION C ER TIFIC
ATE ...