Informações sobre o equipamento:

Cosmed Quark b² Espirômetro
Fabricante:
Cosmed
Modelo: Quark b²
Data: 2003
Categoria: Equipamento Médico e Biomédico
Grupo: Espirômetro
Descrição:
Informações:
                                    Quark b2 is an electrical medical device designed to perform
pulmonary function tests. It is to be used by physicians or
by trained personnel on a physician responsibility.

Main features

The Quark b2 unit contains the main items (analysers,
electronic controls, valves...). It is
connected to the PC through an RS232 interface. The modular
architecture allows to replace the boards without any other
technical service. The unit can be used on a desktop or on
the optional trolley, ideal for the Quark b2, the
cylinders, the PC, the Printer...

The system uses a bi-directional digital turbine. It opposes
a very low resistance to flow (<0,7 cmH2O/l/s to 12 l/s).
The air passing through the helical conveyors, takes a
spiral motion which causes the rotation of the turbine
rotor. The rolling blade interrupts the infrared light
beamed by the three diodes of the optoelectronic reader.
Every interruption represents 1/6 turn of the rotor, this
allows to measure the number of turn in the time.

The O2 and CO2 analysers are temperature-controlled and the
internal pressure and expired flow are monitored for an
higher reliability if the measurements. 
The Quark b2 uses Nafion Permapure ® which is a
semipermeable capillary tube capable of removing the
humidity in excess without altering the gas concentrations..
The analysers calibration is automatic and shows both
graphically and numerically the flow and concentration
signals and the accuracy of the baseline/gain.
measurements.

The Barometric pressure is measured by an electronic
pressure sensor inside the unit. It has a measuring range
between 400 mmHg and 800 mmHg. Ambient temperature and
relative humidity are measured by two sensors contained in
the RH/TA probe. The above measurements are used to
calculate BTPS and STPD correction factors.

Estes manuais estão disponíveis para os equipamentos acima:

Tipo manual: Manual do Usuário Cosmed-10499-Manual-Page-1-Picture
Páginas: 139
Tamanho: 4.22 Mbytes (4427721 Bytes)
Idioma: Inglês
Revisão: XVIII
Manual-ID/Number: C00827-02-91
Data: 01 Janeiro 2003
Qualidade: Documento eletrônico, digitalização, bem legível.
Envie data:
MD5: 81da540db4130f8369b6417cccdf9dab
Downloads: 60 desde 02 Março 2018
Informações:
Getting started
 12
Important
notices ... 
13
Intended use
 ... 
13
Warnings
 ... 
13
Contraindication
 ... 
15
Contraindications for the Spirometer tests
 ...  15
Absolute
contraindications ... 
15
Relative
contraindications ... 
15
Contraindications for Bronchial provocation tests
 ...  15
Absolute
contraindications ... 
15
Relative
contraindications ... 
15
Contraindications for Exercise testing
 ...  15
Environmental condition of use
 ...  16
EMC
 ... 
17
Safety and conformity
 ... 
20
Safety
 ... 
20
EMC ... 
20
Quality Assurance
 ...  20
Medical Device Directive (CE mark)
 ...  20
Keynotes
 ... 
21
Typographic keynotes
 ... 
21
Graphic
keynotes ... 
21
System overview
 ... 
22
Main
features ... 
22
Flowmeter
 ... 
22
Gas
analysers ... 
22
Environmental conditions
measurements ...  22
Before starting
 ... 
23
Checking the packing contents
 ...  23
Quark b2 standard packaging
 ...  23
Warranty registration
 ... 
24
Register the product via software
 ...  24
How to contact Cosmed
 ...  24
Complain, feedback and suggestions
 ...  24
Options/Accessories
 ... 
25
PC configuration required
 ... 
26
Technical features
 ... 
27
Flowmeter
 ... 
27
Oxygen Sensor (O2)
 ...  27
Carbon Dioxide Sensor (CO2)
 ...  27
Humidity absorber
 ...  27
Power Supply
 ... 
27
Environmental Sensors
 ...  27
Dimension and Weight
 ...  27
Aux
inputs ... 
27
28
Pulmonary function tests and measured parameters
 ... 29
Breath by Breath exercise
testing ...  29
Indirect Calorimetry
 ... 
29
Lactate Threshold
(V-Slope) ...  29
O2 Kinetics
 ... 
30
Spirometry Tests (option)
 ...  30
FVC - Forced Vital Capacity
 ...  30
VC/IVC - Slow Vital Capacity and Ventilatory pattern 30
MVV - Maximum Voluntary Ventilation  ...  31
Bronchoprovocation Response
 ...  31
32
Unpacking the system
 ... 33
Setting up the system
 ... 34
Calibration Gas Cylinder
 ...  34
Gas pressure adjustment
procedure ...  34
Connecting cables and
tubing ... 35
Cables and Tubing installation sequence
 ...  36
Assemble the turbine
 ... 37
Assemble the mask and the flowmeter
 ...  37
Using the "Ultimate Seal"
 ...  37
Apply the seal to the mask
 ...  38
To remove seal on mask
 ...  38
Assembling the flowmeter for spirometry tests
 ...  38
Cardiac Output Module
 ...  39
Connecting PC
 ... 
39
Software installation
 ... 40
Installing the software
 ... 
40
Run the software
 ... 
40
PC port configuration
 ... 
40
Software main features
 ... 41
Display
 ... 
41
Tool bar
 ... 
41
Show/hide the toolbar
 ...  41
Dialog windows
 ... 
41
Use of the keyboard
 ...  41
Use of the mouse
 ...  41
Scroll bars
 ... 
41
On line help
 ... 
41
Software version
 ... 
41
Calibrating sensors
 ... 
43
Running the Calibration
program ...  43
Log file
 ... 
43
Setting reference values
 ...  43
To set the reference
values ...  43
Calibrating analysers
 ... 
44
Print the calibration report
 ...  44
Edit the calibration factors
 ...  44
Turbine calibration
 ... 
45
The calibration syringe
 ...  45
Turbine calibration for ergospirometry tests
 ...  45
Assembling the flowmeter
 ...  45
Calibrating the turbine
 ...  46
Turbine calibration for the RMR test
 ...  47
Assembling the flowmeter
 ...  47
Calibrate the
turbine ... 
47
Checking the system signals
 ...  48
The control panel
 ... 
48
Using the control panel
 ...  48
Database Management
 50
Exercise testing patient's database
 ...  51
Enter a new patient
 ... 
51
Find a patient
 ... 
51
Edit patient data
 ... 
51
Delete a patient
 ... 
51
Archive maintenance
 ... 
52
Reorganise the archive
 ...  52
Delete the
archive ... 
52
Delete a test
 ... 
52
Backup and restore
 ... 
52
Backup
 ... 
52
Restore
 ... 
52
Spirometry patient's
database ... 
53
Patient
Card ... 
53
Visit Card
 ... 
53
Test Card
 ... 
54
Import/export a Tests card
 ...  54
Diagnosis Database
 ... 
54
Spirometry program
settings ... 
55
Graphs
 ... 
55
Serial
port ... 
55
Units of
measurements ... 
55
Using extra fields
 ...  55
Customise the fields
 ...  55
Recommendations for the exercise testing
 ... 57
The evaluation of the cardiorespiratory function
 ...  57
Precautions
 ... 
57
Laboratory
 ... 
57
Ending the test
 ... 
57
Preparing the patient
 ... 
57
Before testing
 ... 
57
Patient assent
 ... 
58
Ending the test
 ... 
58
Start Testing
 ... 59
Start a test
 ... 
59
Abort the test without saving data
 ...  59
End the test saving data
 ...  59
View data in real-time
 ... 
59
View graphs in real-time
 ...  60
Parameters to view
 ... 
60
Manual protocol
 ... 
60
Enter Load and Phase
 ...  60
Set the markers
 ... 
60
Automatic protocol
 ... 
60
Modify the load during the test
 ...  60
Set the BPM alarm
 ... 
60
Enter the BPM
 ... 
60
Data management
 ... 61
Viewing data
 ... 
61
View data in table form
 ...  61
Creating graphs
 ... 
61
View data in graph form
 ...  61
Customise the graphs
 ...  62
Switch from graph to data and vice versa
 ...  62
Viewing predicted values
 ...  63
View predicted
values ...  63
Anaerobic (Lactate) Threshold
detection ...  63
View the Lactate Threshold
 ...  63
Detect the Lactate Threshold
 ...  63
Customise graphs for the LT
viewing ...  63
Fittings
 ... 
64
Fit a graph with a linear
regression ...  64
Fit a graph with a Mono-exponential regression  ...  64
Calculate the "Mean Value"
 ...  65
Oxygen Kinetic
 ... 
65
Run the O2 Kinetic function
 ...  65
Information about the Test
 ...  66
View the Information
 ...  66
Modify the
information ...  66
Summary
 ... 
66
View the summary
 ...  66
Print the data
 ... 
66
Print the current window
 ...  67
Print the report
 ... 
67
View the report
 ... 
67
Data Editing
 ... 
68
Editing values and input numerical values
 ...  68
Data filtering
 ... 
68
Using the User fields
 ...  69
Deleting
steps ... 
69
Advanced Editing
 ...  69
Restore the original test
 ...  70
Overwrite the original test
 ...  70
Customise the
desktop ... 
70
Customise the display colours
 ...  70
Smart edit
 ... 
70
Apply the graphical noise suppression
 ...  70
Apply the threshold noise suppression
 ...  70
Customise the parameters
 ...  71
Create a new
parameter ...  71
Create a new predicted parameter
 ...  71
Exporting data
 ... 
72
Export a test
 ... 
72
DDE with Excel
 ...  72
Creating Test Protocols
 ... 
73
Create a new protocol
 ...  73
Software configuration
 ... 
74
Data viewing
 ... 
74
Select the parameters to view
 ...  74
Select the parameters to view during the test
 ...  74
Sort the parameters
 ...  74
Steady State
 ... 
74
Customise the Steady State detection criteria
 ...  74
Printout
reports ... 
75
Set up the printout
 ...  75
Select parameters to be printed
 ...  75
Customise the printout header
 ...  75
Electronic reports (*.pdf)
 ...  76
Print the current window
 ...  76
Print the customised report
 ...  76
Events management during exercise testing
 ...  77
Cardiac Output (option)
 ...  77
Setting up the Quark b2 for the cardiac output
 ...  77
Performing a Cardiac Output test
 ...  78
Flow Volume loops
 ... 
78
Flow Volume loop during the test
 ...  78
O2, CO2 vs Time
 ... 
79
O2, CO2 vs Time during the test
 ...  79
O2 Saturation (optional)
 ...  79
O2 Saturation during the test
 ...  79
Spirogram ... 
79
Spirogram during the test
 ...  79
View the events after the test
 ...  80
Raw
data ... 
80
Save Raw
data ... 
80
Resting Metabolic Rate Test
 82
Metabolism ... 83
Total Metabolic
Rate ... 
83
Resting Metabolic Rate (RMR)
 ...  83
Importance to measure RMR
 ...  83
How to perform a RMR
test ...  83
Recommendations
 ... 84
Resting metabolic rate test using the face
mask ...  84
Resting metabolic rate test using the canopy option
 ...  84
Performing a test using the face mask
 ... 85
Calibrations
 ... 
85
How to prepare a patient
 ...  85
Start the
test ... 
85
Viewing the test
 ... 
87
How to modify the average interval
 ...  87
Print ... 
88
Performing a test using the canopy
option ... 89
Calibrations
 ... 
89
How to prepare the canopy and the patient
 ...  89
Replacement of the power plug
 ...  89
Connecting the
Canopy ...  89
How to prepare the
patient ...  90
Performing the test
 ... 
90
Viewing the test
 ... 
91
How to modify the average interval
 ...  91
Print ... 
91
Sub-maximal Exercise Testing
 92
Introduction
 ... 93
Pre-test screening
 ... 
93
Sub-maximal exercise testing
 ... 94
Considerations with sub-maximal exercise testing
 ...  94
Staffing
 ... 
95
Test termination
 ... 
95
Considerations for accuracy
 ... 96
Performing the test
 ... 97
An example of testing protocol
 ...  97
98
The mixing chamber
 ... 
99
Overview
 ... 
99
Preparing the mixing chamber for a test
 ...  99
Two-way non rebreathing valve description
 ...  99
Patient’s
preparation ... 
100
Performing the test
 ... 
100
102
Setting spirometry
options ... 
103
Spirometry
 ... 
103
Automatic Interpretation
 ...  103
Quality control
 ...  103
Parameters manager
 ... 
104
Predicted values manager
 ...  104
Predicteds set
 ... 
104
Set the current predicted
 ...  105
Formula definition
 ...  105
Page set-up
 ... 
106
Spirometry tests
 ... 
107
Forced Vital Capacity (pre)
 ...  108
Recommendations
 ... 
108
Perform a FVC (pre) test
 ...  108
Test encouragement
 ... 
108
Perform the FVC test with the encouragement  ...  109
Slow Vital Capacity
 ... 
110
Perform a SVC test
 ... 
110
Maximum Voluntary
Ventilation ... 
111
Perform a MVV test
 ... 
111
Bronchial Provocation Test
 ...  112
Bronchodilator test
 ...  112
Perform the test
 ... 
113
Bronchial Provocation protocols Database
 ...  113
Enter a new Bronchial provocation protocol in the
archive ... 
113
Viewing results
 ... 
114
Tests of the current
patient ...  114
Delete a test
 ... 
114
Printing
results ... 
115
Printing Reports
 ... 
115
Printing the active
window ...  115
To print the active window
 ...  115
Printing a series of
reports ...  115
Electronic reports (*.pdf)
 ...  115
Export data
 ... 
116
Export a test
 ... 
116
Systemmaintenance
System maintenance118
 ... 119
Cleaning and disinfection
 ...  119
Preparing the disinfecting solution
 ...  119
Cleaning the turbine
flowmeter ...  120
Precautions during the cleaning of the turbine  ...  120
Masks cleaning and disinfection
 ...  120
Disassembling the different parts of the mask  ...  120
Cleaning the
mask ...  120
Disinfecting the mask
 ...  120
Canopy bubblehood (option) cleaning
 ...  121
RMR reader (option) cleaning
 ...  121
Precautions during the cleaning of the turbine  ...  121
Two-way non rebreathing valve cleaning (option)
 ...  121
Mixing chamber cleaning and disinfection (option)
 ...  121
Permapure maintenance
 ...  121
Inspections
 ... 
122
Replace the
fuses ... 
122
Select the proper power supply voltage
 ...  123
Appendix
 124
Conformity declaration
 ... 125
Service - Warranty
 ... 126
Warranty and limitation of liability
 ...  126
Return goods policy for warranty or non warranty
repair ...  126
Repair Service Policy
 ...  127
Privacy Information
 ... 128
Personal data treatment and purposes
 ...  128
How your personal data are
treated ... 
 ...  128
Holder of the treatment
 ...  128
Customer rights
 ... 
128
Waste of electrical and electronic equipment
 ... 129
Analog - Digital auxiliary inputs
 ... 130
AUX RS232
 ... 
130
Pin-out
assignment ... 
130
Handshaking protocol configuration
 ...  130
AUX1/ AUX4 analog
Inputs ...  130
Enable the AUX input
 ...  130
Converting factors configuration
 ... 132
Pin-out
assignment ... 
132
Heart Rate – TTL input
 ... 133
Calculations references
 ... 134
VO2 and VCO2
 ... 
134
Anaerobic threshold (modified V-Slope)
 ...  134
O2 kinetics
 ... 
134
Cardiac Output
 ... 
134
ATS 94 recommendations
 ...  135
ATS recommendations
 ...  135
Predicted values
 ... 
136
Automatic diagnosis
(algorithm) ...  137
Quality Control Messages
 ...  137
References ... 
139
Gas Exchange References
 ...  139
Indirect
calorimetry ... 
139
Cardiac Output
 ...  139
Spirometry
 ... 
139
Sub-maximal testing
 ...  139
Cosmed-10499-Manual-Page-1-Picture


Comments (0)

No comments yet. Be the first!

Add Comment

* Required information
1000
Captcha Image
Powered by Commentics