The Hewlett-Packard Model 740B is a precision multifunction
instrument that operates as a dc standard Voltage Source a
dc differential voltmeter, a high impedance dc voltmeter and
a dc power and voltage Amplifier The instrument is designed
for use in both the standards laboratory and the field.
1-5. DC STANDARD.
1-6. When used as a dc standard, the Model 740B provides
output voltages from 0 to 1000 V in 4 decade ranges: 1 V, 10
V, 100 V and 1000 V. The output voltage on each range can be
set with 6-digit resolution.
The output voltage is accurate to within ±(0.002% of setting
+ 0.0004% of range). Output voltage is indicated on digital
readout tubes (first five digits) and the meter (sixth
digit). Maximum output current is 50 mA on the 1 V range
decreasing linearly to 20 mA at full output voltage on the
1000 V range. The front panel CURRENT LIMIT control allows
adjustment of the maximum output current from approximately
10% to 100% of the available output current at each voltage
setting. Remote sensing allows the output voltage to be
regulated at the load, eliminating the effects of lead
resistance. The output circuit is floating and guarded. A
pushbutton switch allows the output voltage to be switched
on and off without resetting the voltage dials.
1-7. DIFFERENTIAL VOLTMETER.
1-8. Used as a differential voltmeter, the Model 740B measures dc voltage on seven decade ranges from 1 mVto 1000
V. Voltage measurements are accurate within ± (0.005% of
reading + 0.0004% of range + 1 /xV). The measured voltage is
indicated on five digital readout tubes (first five digits)
and the meter (sixth digit). High input resistance (>10*0 Q
on all ranges above 10 mV) is maintained regardless of
whether or not the voltage dials are nulled. The input
circuit is floating and guarded, allowing accurate
measurements to be made under conditions where ground loops
are a problem.
1-9. HIGH IMPEDANCE VOLTMETER.
1-10. The Model 740B can also be used as a ± 2% floating and
guarded dc voltmeter with 10 voltage ranges from 1 jiV to
1000 V. Input resistance is > lO1^ £2 on all ranges above 10
mV. The measured voltage is indicated on the meter.
1-11 Amplifier 1-12. The Model 740B can be used as a dc power Amplifier in
the Voltmeter and Differential Voltmeter modes of operation
by connecting the source to the input terminals and taking
the output from the output terminals. Voltage gain is unity
on the 1 V range and higher, but increases in 20 dB steps on
the lower ranges to a maximum of 60 dB on the 1 mV range.
Output current characteristics are the same as the DC
Standard mode and the high input resistance characteristics
of the Voltmeter and Differential Voltmeter modes are
retained. Voltage gain accuracy is ±(0.01% + 5 ppm of range
+ 2 uV).
1-13. The Model 740B can also be used as a voltage Amplifier in the Voltmeter and Differential Voltmeter modes with up to
120 dB of voltage gain available at the rear panel RECORDER
OUTPUT terminals. Voltage at these terminals is directly
proportional to meter deflection and is 0 to ± 1 V on all
ranges, 1 µV to 1000 V.
2 Manuais
Manual de serviço e do usuário
Tipo manual:
Manual de serviço e do usuário
Páginas:
33
Tamanho:
3,3 MB
Idioma:
inglês
Revisão:
Manual-ID:
01794-1
Data:
Julho 1972
Qualidade:
Documento digitalizado, tudo legível.
Data de upload:
30 de Dezembro de 2017
MD5:
dec3e515-048c-dd5e-42e6-0ebd18a01a22
Downloads:
497
Download
hewlett-packard--740b--service-user--ID10079.pdf
Manual de serviço e do usuário
Tipo manual:
Manual de serviço e do usuário
Páginas:
113
Tamanho:
12,6 MB
Idioma:
inglês
Revisão:
Manual-ID:
01794-1
Data:
Agosto 1967
Qualidade:
Documento digitalizado, tudo legível.
Data de upload:
19 de Janeiro de 2023
MD5:
c1a4b06b-a2a2-44db-cf96-3d9b516dd5f7
Downloads:
177
Download
hewlett-packard--740b--service-user--ID13867.pdf