The Model 1124A Probe provides high voltage, high impedance,
general-purpose probing capabilities for 50-ohm input, high
frequency instruments. The Probe input RC, 10 megohms
shunted by 10 picofarads, prevents circuit loading while a
FET (field-effect transistor) circuit provides a bandwidth
of dc to 100 MHz. A wide dynamic range, provided in two
ranges (X10, X100), allows direct measurements up to 100
volts with either ac or dc coupling. The maximum safe input
for the X10 range in the dc-coupled mode is ±300 volts (dc
plus peak ac) for frequencies below 100 MHz. The X100 range
extends the maximum safe input to ±500 volts (dc plus peak
ac) for frequencies below 100 MHz. The maximum safe input in
the ac-coupled mode is the same as for the dc-coupled mode
except the dc component of the input signal must not exceed
±200 volts.
Figure 1. Model 1124A Probe
3. The Model 1124A Probe extends the usefulness of 50-ohm
systems such as Hewlett-Packard 183A Oscilloscope systems or
the Hewlett-Packard Model 8553B/8552A Spectrum Analyzer
system by adding general-purpose
Typ ręczny:
Serwis i User Manual
Strony:
10
Rozmiar:
514.58 Kbytes (526928 Bytes)
Język:
english
Rewizja:
Ręczny ID:
01124-90904
Data:
Jakość:
Skanowany dokument, czytając częściowo źle, nie częściowo czytelny.
Data przesłania:
2017 12 30
MD5:
d864d78d88a35cabb1716db81c7f0e69
Pobrane:
446