ELECTRONIC LOAD
Informazioni
Series PLZ-W Electronic Loads can be operated in constant
resistance mode, constant current mode and switching action
electronically, replacing the conventional sliding resistors
(rheostats). Major features of the PLZ-W series are as follows:
1. Digital Meters.
A digital voltmeter and an ammeter with an autorange o£ 3
1/2 digits.
2. A 10-Turn Multi-Dial.
Current values and resistance values may be set very
precisely with the 10 turn multi-dial having a lock mechanists.
3. Function of Presetting.
Current and resistance values may be preset.
4. Built-in Oscillator for Switching Action.
To be used in the constant current mode. Switching action
occurs within the frequency range of 10 Hz - 1 kHz,
5. Display of Loading Power.
Supplied power value is displayed in a digital figure and in
the form of luminescent bar graph.
6. Remote Control Functions.
When operated in the constant current mode the instrument
may be remotely controlled by either an external voltage or
an external resistance. The instrument may also be remotely
controlled by an external resistance when operated in the
constant resistance mode.
Constant voltage mode: 5 - 60 V DC
Constant Current Characteristics.
Stability: 0.1% + 6 mA for loading voltage variation from 4
to 60 V
0.1% + 6 mA for line voltage variation by ±10% Ripple and
noise: 5 mA RMS
Temperature coefficient: 0.02%/°C (standard value)
Rise/fall time: 100 ysec or less (when an internally
installed Oscillator is used).
Constant Resistance Characteristics.
Stability: ±10% loading voltage variation:
(Stability) (Input Voltage (Setting resistance)
Variation)
4 - 10 V 1 Ω
4 - 40 V 10 Ω
Stability ±10% variation of line voltage: 0.1% + 6 mA
Temperature coefficient: 0.03%/°C standard value (at the
minimum resistance value of a range).
Remote Control.
Constant current: External resistance 0 - 1000 Ω
External voltage 0 - 10 V DC Constant resistance: External
resistance 0 - 1000 Ω Constant voltage: External voltage 0 -
10 V DC
Protecting Functions.
Overvoltage protection: Approx. 65 V DC Overcurrent
protection: Approx. 65 A DC Overpower protection: Approx. 330 W
Wrong polarity input protection: By use of a diode
Overheating protection: 100°C ±5°C at cooling package Line
input fuse rating: 1 A
Tipo di manuale:
Servizio e manuale utente
Pagine:
51
Dimensione:
1.91 Mbytes (2001157 Bytes)
Lingua:
english
Revisione:
ID del manuale:
Data:
Qualità:
Documento scansionato, la lettura in parte male, non in parte leggibile.
Data caricamento:
2015 08 14
MD5:
302d01642b3ec579a7cff6c74d402c2a
Scaricamenti:
575
Informazioni
SECTION 1. GENERAL ...
1—1. Introduction
1-2. Specifications
SECTION 2. OPERATION ...
2-1. Notes Before Operation
...
2-2. Front and Rear Panel Descriptions
2-3. Initial Setting
...
2-4. Operation in the Constant Resistance Mode 2-5. Use of
"xlO" Range in the Constant Resistance
Mode
...
2-6. Graph of Resistance Range and Power Limit
Relation
...
2-7. Operation in the Constant Current Mode ...
2-8. Operation in the Switching Mode ...
2-9. Operation in the Constant Voltage Mode ...
2-10. Indication of Loading Power
2-11. Presetting the Power Limit ...
2-12. Remote Control with an External Resistance (Constant
Resistance or Constant Current}
2-13. Remote Control with an External Voltage (Constant
Current only)
2-14. Remote Control of Load Switch by an External
Terminal
...
2-15. One-control Parallel Operation ...
SECTION 3. OPERATING PRINCIPLES
3-1. Constant Current Operation
3-2. Constant Resistance Operation
3-3. Block Diagram ...
SECTION 4. MAINTENANCE
...
4-1. How to Open the Chassis Case
...
4-2. Inspections and Cleaning
...
4-3. Calibration
...
WARRANTY
...