Keithley 2425 Metro di livello
Produttore:
Nessuna immagine disponibile!
Forse puoi caricare una foto per la
Keithley 2425?
Se avete altri manuali per la
Keithley 2425 potete caricare i file qui .
Modello:
2425
Data:
1998
Categoria:
Gruppo:
Descrizione:
SourceMeter
Informazioni
Product overview
The SourceMeter combines a precise, low-noise, highly stable
DC Power supply with a
low-noise, highly repeatable, high-impedance multimeter. It
has 0.012% basic accuracy
with 51⁄2-digit resolution. At 51⁄2 digits, the SourceMeter
delivers 520 readings/second
over the IEEE-488 bus. At 41⁄2 digits, it can read up to
2000 readings/second into its inter-
nal buffer. The unit has broad source and measurement ranges:
Model(2400:
•
Source voltage from 5μV to 210V; measure voltage from 1μV
to 211V.
•
Source current from 50pA to 1.05A; measure current from
10pA to 1.055A.
•
Measure resistance from 100μΩ (<100μΩ in manual ohms) to 211MΩ.
•
Maximum source power is 22W.
Model(2410:
•
Source voltage from 5μV to 1100V; measure voltage from 1μV
to 1100V.
•
Source current from 50pA to 1.05A; measure current from
10pA to 1.055A.
•
Measure resistance from 100μΩ (<100μΩ in manual ohms) to 211MΩ.
•
Maximum source power is 22W.
Model(2420:
•
Source voltage from 5μV to 63V; measure voltage from 1μV to
63.3V.
•
Source current from 500pA to 3.15A; measure current from
100pA to 3.165A.
•
Measure resistance from 10μΩ (<10μΩ in manual ohms) to 21.1MΩ.
•
Maximum source power is 66W.
Models 2425 and(2430:
•
Source DC or pulse voltage from 5μV to 105V; measure
voltage from 1μV to
105.5V.
•
Source DC current from 500pA to 3.15A; measure DC current
from 100pA to
3.165A.
•
Source pulse current from 500pA to 10.5A; measure pulse
current from 100pA to
10.55A. (Model 2430 only.)
•
Measure resistance from 10μΩ (<10μΩ in manual ohms) to 21.1MΩ.
•
Maximum DC source power is 110W.
•
Maximum pulse source power is 1.1kW. (Model 2430 only.)
Model(2440:
•
Source voltage from 5μV to 42V; measure voltage from 1μV to
42V.
•
Source current from 500pA to 5.25A; measure current from
100pA to 5.25A.
•
Measure resistance from 10μΩ (<10μΩ in manual ohms) to 21.1MΩ.
•
Maximum source power is 66W.
NOTESThe Models 2400
2410
2420
2425
2430
and 2440 are
Y2K compliant.
Models with a “-C” suffix have the Contact Check option. See
Appendix F.
Some additional capabilities of the SourceMeter include:
Concurrent measurements of all three functions over the
remote Interface.
Source-measure sweep capabilities (linear and logarithmic
staircase sweeps, source sweep list of up to 2500 points,
memory sweep of up to 100 instrument setups).
6-wire Ω measurement with programmable I-source or V-source
with V or I clamp.
4-quadrant source and sink operation.
Up to 12 stages of limit testing with a built-in comparator
for pass/fail testing.
Digital I/O for stand-alone binning operations or Interface
to component handler.
Programming language and remote interfaces — The SourceMeter
uses the SCPI programming language and two remote Interface
ports (IEEE-488/GPIB and RS-232C).
Trigger-Link Interface to Keithley Series 7000 switching
hardware.
Math expressions — Five built-in, up to five user-defined
(bus only).
Reading and setup storage — Up to 2500 readings and seven
setups (five user defaults, factory default, *RST default)
can be stored and recalled.
Closed-cover calibration — The instrument can be calibrated
either from the front panel or remote Interface.
Tipo di manuale:
Manuale d'uso
Pagine:
592
Dimensione:
2.49 Mbytes (2614077 Bytes)
Lingua:
english
Revisione:
G
ID del manuale:
2400S-900-01
Data:
2002 05 01
Qualità:
Documento elettronico, senza ricerca, molto ben leggibile.
Data caricamento:
2018 08 23
MD5:
2a4420abc1106276b376316f115a7fa0
Scaricamenti:
2804
Tipo di manuale:
Manuale d'uso
Pagine:
53
Dimensione:
723.99 Kbytes (741362 Bytes)
Lingua:
english
Revisione:
C
ID del manuale:
2400S-903-01
Data:
Qualità:
Documento elettronico, senza ricerca, molto ben leggibile.
Data caricamento:
2014 03 07
MD5:
3c9ff54e64ca0a36a6a083410e1d09e4
Scaricamenti:
3710
Informazioni
Quick Results Guide
This guide is designed to familiarize users with fundamental
operation (front panel and
remote) of the Keithley 2400 Series SourceMeters. For
comprehensive information on all
aspects of SourceMeter operation, refer to the 2400 Series
SourceMeter User’s Manual.
Operation information in this guide is divided into four
parts; (1) Fundamental sourcemeasure
operations, (2) Settings to optimize performance, (3)
Features to enhance DUT testing
and (4) More testing techniques. This format allows a new
user to easily progress from basic
simple operation to more complex procedures.