The Model 8502A Digital multimeter is a 6 1/2 digit
instrument employing microprocessor control and a bus
structure. Memory programming either from the front panel or
through a remote Interface permits a number of operations to
be performed on the measured input before it is displayed.
The standard configuration allows for measurement of dc
volts in 5 ranges. Four ranges are available for ac volts
when either optional ac Converter is installed. Resistance
can be measured in 8 ranges. Current can be measured in 5
ranges.
Modular Construction
Considerable versatility is realized through the 8502A’s
unique construction. All active components are contained in
modules which plug into a mainframe motherboard. This
module-motherboard mating, combined with bus architecture
and microprocessor control, yields both ease of option
selection and reduced downtime.
Microprocessor Control
All modules function under direct control of a
microprocessor based controller. Each module is addressed by
the controller as a memory location. External reference
values and offsets can be applied separately, stored in
memory, and automatically used as factors in all subsequent
readings. Digital filtering utilizes averaged samples for
each reading.
Recirculating Remainder A/D Conversion
The 8502A adapts Fluke’s patented recirculating remainder
(R2) A/D conversion technique to microprocessor control.
This combination provides fast, accurate, linear
measurements and long-term stability.
2 Manuels
Manuel d'entretien et d'utilisation
Type manuel:
Manuel d'entretien et d'utilisation
Pages:
378
Taille:
31,9 Mio
Langue:
anglais
Révision:
Manuel-ID:
471466
Date:
juillet 1981
Qualité:
Document scanné, tout est lisible.
Date de téléchargement:
14 janvier 2018
MD5:
639cbce3-8c75-f0c9-dea9-faffad128c4a
Information
This manual comprises eight modular sections. You will find
herein up-to-date information for installing, operating and
maintaining the Fluke Model 8502A Digital multimeter Generally, complete descriptions and instructions are
provided for the 8502A mainframe, modules necessary for DC
Volts and DC Ratio measurement, and for any optional modules
(AC Converter Ohms Converter etc.) that you may have
ordered with your 8502A Specifically, each section contains:
1. Section 1 — General description, specifications.
2. Section 2 — Operating instruction, capabilities.
3. Section 3 — Theory of operation, including simplified
schematic and functional block diagrams.
4. Section 4 — Maintenance, adjustments and troubleshooting.
5. Section 5 — List of replaceable parts, with parts locators.
6. Section 6 — Option and accessory information, including
complete information on any option ordered with the 8502A As the need arises to broaden your8502A’s capabilities, the
most recent information will be included with any options
you order.
7. Section 7 — General information (list of abbreviations,
federal supply codes, Service Centers, and Sales
Respresentatives).
8. Section 8 - Schematic diagrams.
Téléchargements:
964
Téléchargements du mois dernier:
3
Télécharger
fluke--8502a--service-user--ID10110.pdf
Supplément au manuel d'entretien et d'utilisation
Type manuel:
Supplément au manuel d'entretien et d'utilisation
Pages:
64
Taille:
5,4 Mio
Langue:
anglais
Révision:
Manuel-ID:
577049
Date:
juillet 1981
Qualité:
Document scanné, tout est lisible.
Date de téléchargement:
14 janvier 2018
MD5:
04b4a2ab-0b6b-8fba-26b8-d422f514c75d
Information
This addendum modifies the standard 8502A Instruction Manual
to cover operations and servicing of the 8502A AT Digital
Multimeter. If no reference is made to a numbered paragraph,
figure, or table, that item in the 8502A manual remains
correct, except for the change in nomenclature.