Elenco Precision F-1000 Counter
Manufacturer:
Image 1 of 1
If you have any other photos or manuals for the
Elenco Precision F-1000 you can upload the files here .
Model:
F-1000
Date:
0
Category:
Group:
Description:
1GHz Multifunction Counter
Information
INTRODUCTION
The Elenco model F-1000 is a 10Hz to 1000MHz
multiple-function Counter that features a temperature
control crystal oven for high accurate readings. Other
features are eight high brightness seven segment
LED display, low power consumption circuit design and full
input signal conditioning. The F-1000 is also
small in size and light weight for easy portability.
The model F-1000 has four functions in one. It is a 1000MHz
frequency Counter with accuracy of
,3Hz/million per month. This accuracy is achieved because of
the temperature control crystal oven. It
also has a self-test function which allows checking the
output for proper operation. The third function is a
period measurement which allows you to measure time interval
between pulses. The fourth function is a
totalizer Counter which allows you to count pulses. All
these functions are accomplished by a single LSI
integrated circuit. The input signal is AC or DC coupled and
can be conditioned by attenuation of the
signal.
The location of controls, indicators, connectors and other
information on this model is provided in this
manual. It is recommended that you read and understand all
information in this manual before attempting
to operate this instrument.
SPECIFICATIONS:
MEASURING MODE
Frequency measurements
CHANNEL A.
□ Range: 10Hz to 10MHz direct Counter
10MHz to 100MHz prescaled by 10.
□ Resolution: Direct counter: 1, 10, 100Hz switch selectable.
Prescaled: 10, 100, 1000Hz switch selectable.
□ Gate Time: 0.01S, 0.1S, 1S switch selectable.
□ Accuracy: ± 1 count ± time base error.
CHANNEL B.
□ Range: 100MHz to 1GHz.
□ Resolution: 100Hz, 1KHz, 10KHz switch selectable.
□ Gate Time: 0.027S, 0.27S, 2.7S switch selectable.
□ Accuracy: ± 1 count ± time base error.
Period Measurements (Channel A)
□ Range: 10Hz to 2.5MHz.
□ Resolution: 10-7S, 10-8S, 10'9S switch selectable.
□ Accuracy: ± 1 count ± time base error.
Totalize Measurements (Channel A)
□ Range: 10Hz to 10MHz.
□ Resolution: ± 1 count of input.
INPUT CHARACTERISTICS
CHANNEL A.
□ Input Sensitivity: 25mV RMS sine wave or 70mV p-p.
□ Attenuation: x 1 , x 20 fixed.
□ Impedance: Approx. 1M ohm less then 35 pF.
□ Maximum voltage: 250V (DC + AC rms). Exceeding this limit
is not recommended.
CHANNEL B.
□ Input Sensitivity: 15mV RMS sine wave or 50mV p-p.
□ Impedance: Approx. 50 ohm.
□ Maximum voltage: 3V. Exceeding this limit is not recommended.
TIME BASE
□ Frequency: 10MHz, 3.90625MHz Temperature Control Oven.
□ Aging Rate: ± 3 x 10'7/monjh.
□ Temperature: ± 1 x 10‘6, 0°C to 40°C.
□ Line Voltage: ± 1 x 10'7 for 10% change.
□ Warm-up time: 20 minutes when cold started at 25°C.
GENERAL
□ Display: 8 digits, 7mm red LED display with decimal point,
gate, overflow, KHz, MHz and uS indication.
□ Check: Counts internal 10MHz time base Oscillator.
□ Power Requirement: Line: 115/230V ± 15%, 45Hz-70Hz.
Internal Battery: Option.
□ Temperature: Rated range of use: - 5°C - + 50°C.
Manual type:
Service and User Manual
Pages:
9
Size:
421.91 Kbytes (432038 Bytes)
Language:
english
Revision:
Manual ID:
Date:
Quality:
Scanned document, all readable.
Upload date:
2015 03 07
MD5:
74890562953bc945c1ec1b779db78c56
Downloads:
942
Information
This is the user manual with schematics & calibration
routine of Multifunction Counter Elenco F-1000,
There are several others brand that use the same model :
Elix 8100A, Tenma 72-640, Protek B-818 and several others
that have the model as 8100A
INTRODUCTION AND SPECIFICATIONS
OPERATION
1. Introduction
2. Preparation For Use
3. Front Panel Features
4. Operating Characteristics
A. Frequency Measurements
B. Period Measurements
C. Totalize Measurements
D. Check Mode
CALIBRATION
1. Introduction
2. Test Instruments Required
3. Time Base Radio Frequency Adjustment
4. Trigger Level Adjustment
SCHEMATIC DIAGRAMS
PARTS LIST