Keithley 194A Digital Multimeter
Hersteller:
Kein Bild verfügbar!
Vielleicht kannst du ein Bild für
Keithley 194A
hochladen ?
Wenn du andere Handbücher für ein
Keithley 194A hast,
kannst du die Dateien hier hochladen .
Modell:
194A
Datum:
1985
Kategorie:
Gruppe:
Beschreibung:
High Speed Voltmeter
Information
FEATURES
The Model 194A High Speed Voltmeter is a high speed DC
voltage sampling instrument suitable for a wide variety of
applications, including the analysis of laboratory
phenomenon, as well as the characterization of transient and
periodic waveforms.
The Model 194A can sample at rates up to 1MHz with 2 1/2
digit, 8-bit usable resolution, or up to 100kHz with 4 1/2
digit, 16-bit resolution. Each A/D channel has 64K bytes of
memory allowing up to 65,535 samples to be stored for later
analysis with one of the many Model 194A mathematical functions.
Key Model 194A features include:
• Autoranging—Autoranging allows the instrument to measure a
wide dynamic range of input signals.
• Zero—A zero feature allows a baseline signal level to be
subtracted from subsequent readings. The baseline can either
be taken from an applied signal, or keyed in from the front
panel or programmed over the IEEE-488 bus.
• Filtering—Internal analog filtering with selectable poles
of 50kHz or 500kHz is available to reduce the effects of noise.
• Math Functions—A number of mathematical functions
including average, integral, standard deviation, true RMS,
peak, and peak-to-peak are included with the instrument
software.
• Dual-channel Operation—Two separate, isolated A/D channels
are available with the optional Model 1944A A/D Module
installed.
• IEEE-488 Interface—A standard feature of the Model 194A
that allows the instrument to be controlled from a computer.
• Real Time Output—This port allows instrument data to be
transmitted to a computer or similar device at the sampling
rate via a user-supplied Interface.
• Programmable Sampling Rates—The instrument can be
programmed to sample data as fast as 1MHz, or as slow as 1Hz.
• Programmable Number of Samples—The number of samples per
measurement sequence can be programmed to any value between
1 and 32,767 (16-bit resolution) or 1 and 65,535 (8-bit
resolution).
• Programmable Triggering—The instrument can be triggered
from the input signal, an external trigger signal, from the
other channel (when the Model 1944A option is installed), or
over the IEEE-488 bus. Programmable triggering parameters
include single or continuous modes, delay, slope, and level.
• XY Mode—Allows the instrument to drive external display
devices such as oscilloscopes and plotters. Internal
Software allows easy generation of graphs from sampled data.
• Translator Mode—Simplifies IEEE-488 programming by
allowing the use of English-like syntax.
• External Sample Control—Allows sample collection to be
synchronized to external events
Dokumenttyp:
Bedienungsanleitung
Seiten:
63
Größe:
946.54 Kbytes (969258 Bytes)
Sprache:
english
Revision:
A
Dokument-ID:
194A-900-01
Datum:
1989 01 01
Qualität:
Gescanntes Dokument, alles ist lesbar.
Upload Datum:
2018 08 23
MD5:
8e05bba89b30c1f80118467bb7f5ac73
Downloads:
1244
Information
Programming Manual
USING THIS GUIDE
This reference and programming guide contains condensed
specifications, a summary of front panel operating controls,
and IEEE-488 programming commands and examples for the
Models 194A
1941
and 1944A, It is intended to provide a
quick reference to the many available features of your
instrument as well as the many programming commands
available to control the instrument over the IEEE-488 bus.
Because of the limited amount space available in this guide,
many details a bout'instrument operation will not be found
here. Refer to the Model 194A Instruction Manual for aspects
of instrument operation not covered in this guide.
This guide describes the following aspects of the Model(194A:
• Front panel operation including XY mode.
• A description of mathematical functions.
• Input and output signal connections.
• IEEE-488 programming information, including commands, data
and status formats, and example programs for typical
controllers.
Dokumenttyp:
Bedienungsanleitung
Seiten:
68
Größe:
2.16 Mbytes (2264777 Bytes)
Sprache:
english
Revision:
A
Dokument-ID:
194A-903-01
Datum:
1987 01 01
Qualität:
Gescanntes Dokument, alles ist lesbar.
Upload Datum:
2018 08 21
MD5:
b027ed520b313320366cacd87271113e
Downloads:
455
Information
IEEE-488 Programming Guide
ABOUT THIS PROGRAMMING GUIDE
This programming guide is intended as a tutorial on a number
of programming concepts associated with the Model 194A, It
is not intended to replace the IEEE-488 programming section
of the Instruction Manual, which should be consulted for
detailed programming information.
Programming concepts discussed in this guide include:
• Programming a single-channel unit and obtaining data.
• Programming a dual-channel unit and requesting data.
• Single/continuous arming and triggering.
• Using binary data transfer to optimize speed.
• Using SRQ and error word to minimize controller overhead.
• Buffer operation, including use of buffer pointers.
• Using Translator to increase readability and emuIate~other
instrumentation.
In addition to these fundamental concepts, applications
programs for external sample triggering, plotting data, and
trigger performance and timing are also included. Finally,
reference information such as device-dependent commands,
data formats, and error words is also supplied.
Dokumenttyp:
Service- und Bedienungsanleitung
Seiten:
296
Größe:
27.05 Mbytes (28364980 Bytes)
Sprache:
english
Revision:
Dokument-ID:
194A-901-01
Datum:
1985 01 01
Qualität:
Gescanntes Dokument, alles ist lesbar.
Upload Datum:
2018 08 21
MD5:
dec40fad073cd55e30823f390d2a08de
Downloads:
691