Calibration Fixture
Information
The 067-0523-00 Calibration Fixture is used with Tektronix
580-Series oscilloscopes. The calibration fixture provides a
reference signal for setting the Oscilloscope vertical
Amplifier gain, a fast-rise voltage step for adjusting
Oscilloscope vertical Amplifier transient response, and a
load bank which loads the Oscilloscope power supplies over
their full current range. The 067-0523-00 allows a check of
the DC balance of the Oscilloscope vertical Amplifier. It
may also be fed external signals which might be used in
calibration procedures.
580-Series Oscilloscope vertical amplifiers that have been
calibrated with the 067-0523-00 have a uniform transient
response. The amplitude calibration signal has a long-term
amplitude stability. The amplitude signal circuit is driven
by the dual-trace alternate sync pulse generated by the
oscilloscope. The dual-trace display shows the presence of
the sync pulse.
Dokumenttyp:
Service- und Bedienungsanleitung
Seiten:
29
Größe:
20.94 Mbytes (21959063 Bytes)
Sprache:
english
Revision:
Dokument-ID:
Datum:
Qualität:
Gescanntes Dokument, alles ist lesbar.
Upload Datum:
2017 12 25
MD5:
ca5e2418be7a6a35c307eed82e585a63
Downloads:
1511
Information
Characteristics
Operating Instructions
Circuit Description
Calibration and Maintenance
Abbreviations and Symbols
Parts Ordering Information
Electrical Parts List Mechanical Parts List & Diagrams