Fluke 8502AT Multímetro digital
Fabricante:
Nenhuma foto disponível!
Talvez você possa carregar uma foto para os
Fluke 8502AT ?
Se você tem outros manuais para os
Fluke 8502AT você pode carregar os arquivos aqui .
Modelo:
8502AT
Data:
1981
Categoria:
Grupo:
Descrição:
Informação
The 8502AT is an 8502A that has been altered to operate at
400 Hz ±5% and has a Thermal True RMS Converter module
installed. In addition the -02 Ohms Converter, -04
Calibration Memory, -05 IEEE Interface, and -08 A Isolator
Options are included as standard. The Thermal True RMS
Converter module allows the operator to measure the true rms
value of an ac signal at accuracies of 50 ppm with a maximum
response time of 6 seconds. This response time compares
favorably with that of existing thermal transfer standards
which can take up to 5 minutes to complete a measurement.
Multimeter operation at 400 Hz results in readings that are
5% slower than the standard 60 Hz unit. The 5% slower
readings are based on timing of 57 Hz which occurs when the
input line frequencies are at 400 Hz ±5%.
Tipo de manual:
Suplemento de Serviço e Manual do Usuário
Páginas:
64
Tamanho:
5.41 Mbytes (5673655 Bytes)
Idioma:
english
Revisão:
ID de manual:
577049
Data:
1981 07 01
Qualidade:
Scaned documento, todas legível.
Data de upload:
2018 01 14
MD5:
04b4a2ab0b6b8fba26b8d422f514c75d
Downloads:
1005
Informação
This addendum modifies the standard 8502A Instruction Manual
to cover operations and servicing of the 8502A.AT Digital
Multimeter. If no reference is made to a numbered paragraph,
figure, or table, that item in the 8502A manual remains
correct, except for the change in nomenclature.