Keithley 515 Ander
Fabrikant:
Afbeelding 1 van 1
Als je nog andere foto's of handleidingen hebt voor de
Keithley 515 kun je de bestanden hier uploaden .
Model:
515
Datum:
1967
Categorie:
Groep:
Omschrijving:
Megohm Bridge
Informatie
The model 515 Megohm Bridge is a Wheatstone Bridge for
measuring resistors from 10⁵ ohms to 10⁻¹⁵ ohms with
accuracies from 0.05% to 1.0%. It is complete, with an
electrometer null detector, shielded enclosure for the
unknown, and a bridge voltage supply.
A unique system of switches is provided to allow corrections
to be made for the slow changes in resistance of the
standard high megohm resistors. This enables all values of
resistance to be read with rated accuracy, directly from the
bridge dials.
Bridge voltages in one volt steps up to 10 volts are
available from the internal supply. With external supplies,
voltages as high as 1,000 volts can be used.
A connector is provided so that unknown resistors can be
measured outside the instrument as well as in the built-in
shielded enclosure.
Handleidingen type:
Service-en gebruikershandleiding
Pagina’s:
46
Grootte:
2.03 Mbytes (2130136 Bytes)
Taal:
english
Herziening:
Handleidingen-ID:
Datum:
Kwaliteit:
Gescande document, lezen deels slecht, deels niet leesbaar.
Upload datum:
2013 01 13
MD5:
356195009ba7e0413669651cc6c9fbbb
Downloads:
852
Handleidingen type:
Service-en gebruikershandleiding
Pagina’s:
46
Grootte:
2.04 Mbytes (2141300 Bytes)
Taal:
english
Herziening:
Handleidingen-ID:
Datum:
1962 01 01
Kwaliteit:
Gescande document, lezen deels slecht, deels niet leesbaar.
Upload datum:
2018 08 23
MD5:
a8eb2a5fe2f3f6b9a7908a7608248c9c
Downloads:
635
Informatie
III-9
1. Null Detector
Sensitivity ... III-9
2. Null Detector Zero
Drift ... 111-10
3. Resolution of the
Readout ... 111-10
4. Accuracy of the Read-
out and Standard
Resistors ... . 111-10
5. Temperature and Voltage
Coefficients in the
Bridge Resistors ... III-I2
6. Leakage Resistance
Across the Unknown, . . 111-12
7. Errors Caused by Guard
to Ground Resistance, . Ill-12
Section Page
8. Time Constants - Slow
Response ... 111-13
9. Transients Caused by
Push-to-Read Switch. , 111-14
10. Verification of the
Accuracy ... 111-14
IV. CIRCUIT DESCRIPTION ... IV-1
A. Power Supply ... IV-1
B. Null Detector ... IV-1
C. Bridge Circuitry ... IV-3
D. Overvoltage Protection. . . IV-4
V. MAINTENANCE ... V-l
6. REPLACEABLE PARTS ... 6-1
6-1. Replaceable Parts List. . 6-1
6-2. How to Order Parts, . . .
6-1 model 515 Replaceable
Parts List ... 6-2
model 515 Schematic
Diagram 14522D ... 6-10