Keithley 7059 Interfaz
Fabricante:
Imagen 1 de 2
Imagen 2 de 2
Si tienes otras fotos o manuales para el
Keithley 7059 puedes subir los archivos aquí .
Modelo:
7059
Fecha:
1989
Categoría:
Grupo:
Descripción:
Low Voltage Scanner Card
Información
The Model 7059 is a low voltage scanner Plug-In card which
is field-installable in the model 705 and model 706 scanner
mainframes. The Model 7059 will switch up to 10 channels.
For low-level transducer and thermocouple output switching
the Model 7059 is designed to introduce a minimum of thermal
voltage error ( < 1µV input to output with copper leads),
and ensure that high isolation (10¹²Ω) is maintained between
input signals. Expected relay life (10⁶ closures) is
obtained when signals less than 10V or 10mA are switched.
However, peak signals up to 200V or 100mA may be switched
with the scanner, Quick disconnect screw terminals are used
to facilitate input and output connections to the scanner
Plug-In card (relay cardl. The Plug-In card is easily
removed through the rear panel of the scanner mainframe.
Type manuel:
Servicio y Manual del usuario
Páginas:
19
Tamaño:
428.73 Kbytes (439022 Bytes)
Idioma:
english
Revisión:
E
Manuel-ID:
7059-901-01
Fecha:
Formato:
Documento Scaned, todo legible.
Fecha de subida:
2014 03 02
MD5:
d4f1f0049da54d8f2ad52b051f2acdc7
Descargas:
653
Información
SECTION 1-GENERAL INFORMATION
1.1 Introduction
... 1
1.2 Warranty Information
... 1
1.3 Manual Addenda
... 1
SECTION 2-OPERATION
2.1 Introduction
... 3
2.2 Wiring and
Installation ... 3
2.3 Operating
Considerations ... 3
2.4 Operating
Hints ... 4
SECTION 3-MAINTENANCE
3.1 Introduction
... 5
3.2 Required Test
Equipment ... 5
3.3 Performance
Verification ... 5
3.3.1 Thermal Offset
Test ... 5
3.3.2 Contact Resistance
Test ... 6
3.3.3 Isolation
Test ... 7
SECTION 4-REPLACEABLE PARTS
4.1 Introduction
... 9
4.2 Replaceable
Parts ... 9
4.3 Ordering
Information ... 9
4.4 Factory
Service ... 9
4.5 Component Layout and Schematic
Diagram ... 9
LIST OF FIGURES
1 Plug-In Card Installation and
Assembly ... 3
2 Howto Use GUARD
Connections ... 4
3 Test Set Up for Thermal Offset Test
... 6
4 Shorting HI and LO of all Channels on
Card ... 6
5 Test Set Up for Contact Resistance
Test ... 7
6 Channel Isolation Test
Setup ... 8
7 GUARD Differential Isolation Test
Setup ... 9
8 Common Mode Isolation Test
Setup ... 9
9 Model 7059 Component Location
Drawing ... 13
10 Model 7059 Schematic Diagram
... 15
Type manuel:
Hoja de datos
Páginas:
1
Tamaño:
483.18 Kbytes (494772 Bytes)
Idioma:
english
Revisión:
Manuel-ID:
Fecha:
Formato:
Documento electrónico, sin análisis, muy bien legible.
Fecha de subida:
2014 03 07
MD5:
e50e59635b9c8400231125434185036b
Descargas:
961