Thurlby Thandar Instruments TTI TSX Series Power supply
Manufacturer:
No picture available!
Maybe you can upload a photo for the
Thurlby Thandar Instruments TTI TSX Series ?
If you have any other manuals for the
Thurlby Thandar Instruments TTI TSX Series you can upload the files here .
Model:
TSX Series
Date:
2009
Category:
Group:
Description:
Programmable High Current DC Power Supplies
Information
OUTPUT SPECIFICATIONS
Output Voltage Range: 0V to 35.3V (35V/10A); 0V to 18.15V
(18V/20A).
Output Current Range: 0.01A to 10.2A (35V/10A); 0.01A to
20.2A (18V/20A).
Output Voltage Setting: Direct keyboard entry or
quasi-analogue rotary control; setting
resolution 10mV.
Output Current Setting: Direct keyboard entry or
quasi-analogue rotary control; setting
resolution 10mA.
Output Mode: Constant voltage or constant current with
automatic crossover.
Output Switch: Electronic. Preset voltage and current
displayed when off.
Output Terminals: 4mm terminals on front panel, screw
terminals at rear.
Sensing: Remote via rear panel screw terminals or direct via
shorting strip
(supplied).
Output Impedance: <1mΩ in constant voltage mode.
>5kΩ in constant current mode (voltage limit at max.).
Output Protection: Forward protection by
Over-Voltage-Protection (OVP) trip;
maximum voltage that should be applied to the terminals is 50V.
Reverse protection by diode clamp for reverse currents up to 3A.
OVP Setting: Direct keyboard entry.
OVP Range: 1V to 40V (35V/10A); 1V to 25V (18V/20A).
OVP Delay: <200μs
Load & Line Regulation: <0.01% of maximum output for 90%
load change or 10% line
change.
Ripple & Noise: <1mV rms typical in constant voltage.
<3mA rms typical in constant current.
HF Common Mode Noise: Typically <3mV rms, <10mV peak.
Transient Response: 20μs to within 50mV of setting for 90%
load change.
Temperature Coefficient: typically <100 ppm/°C.
Protection Functions: Overvoltage trip; regulator
overtemperature; sense miswiring.
Status Indication: Output off lamp; constant voltage mode lamp;
constant current mode lamp; trip message on display.
METER SPECIFICATIONS
Meter Types: Dual 4 digit meters with 12·5mm (0.5") LEDs.
Reading rate 4Hz.
Meter Resolution: 10mV, 10mA
Meter Accuracy: Voltage 0.2% of reading ± 1 digit, current
0.5% of reading ± 1 digit.
KEYBOARD FUNCTIONS
Delta Mode: Increase or decrease voltage or current in
user-selectable steps.
Store/Recall: Store and recall voltage, current and OVP
levels from non-volatile
memory (25 memories).
Interface Selection: Set digital Interface type (RS232 or
GPIB), baud rate and address.
Manual type:
User Manual
Pages:
95
Size:
1.96 Mbytes (2052198 Bytes)
Language:
english, german, french
Revision:
Issue 12
Manual ID:
48511-0130
Date:
2009 01 01
Quality:
Electronic document, no scan, very well readable.
Upload date:
2017 01 14
MD5:
11956a66d324f0e3355b44f4cb497be5
Downloads:
865
Information
Specification 3
Safety 5
EMC 6
Installation 7
Manual Operation 8
Maintenance and Repair 11
Remote Operation 12
ARC Interface 13
Remote Commands 22
Instructions en Francais
Sécurité 34
Montage 35
Fonctionnement manuel 36
Entretien et Réparations 40
Fonctionnement à distance 41
Commandes à distance 52
Bedienungsanleitung auf Deutsch
Sicherheit 64
Installation 65
Handbedienung 66
Wartung und Instandsetzung 70
Fernbetrieb 71
Fernbefehle 82
1
Manual type:
User Manual
Pages:
95
Size:
867.72 Kbytes (888548 Bytes)
Language:
english, german, french
Revision:
Issue 13
Manual ID:
48511-0130
Date:
2013 01 01
Quality:
Electronic document, no scan, very well readable.
Upload date:
2017 01 14
MD5:
1b73ba08af63e918fc8289d61fdc62e3
Downloads:
399
Information
Specification 2
Safety 4
EMC 6
Installation 7
Manual Operation 8
Maintenance and Repair 11
Remote Operation 12
ARC Interface 13
Remote Commands 22
Instructions en Francais
Sécurité 34
Montage 35
Fonctionnement manuel 36
Entretien et Réparations 40
Fonctionnement à distance 41
Commandes à distance 52
Bedienungsanleitung auf Deutsch
Sicherheit 64
Installation 65
Handbedienung 66
Wartung und Instandsetzung 70
Fernbetrieb 71
Fernbefehle 82