Equipment information:

Fluke 5205A Amplifier
Model: 5205A
Date: 1976
Category: Measurement instruments
Group: Amplifier
Description: Precision Power Amplifier
                                    The Model 5205A Precision Power Amplifier is a dc coupled,
broad band inverting amplifier with a set gain of 100. The
5205 A may be used in combination with the 5200A AC
Calibrator for a combined output ranging from 100 (100.000)
to 1100 (1100.000) volts rms at frequencies up to 100 kHz.
When the 5205A is operating as an independent amplifier ac
voltages from 0V to 1100V rms at currents up to 200 mA rms
may be obtained. The waveforms may vary from a large
amplitude sinusoidal to complex waveforms of DC, however,
peak voltage specifications limit the output to 1500 volts.
DC voltages from zero to 1500 volts may be obtained at
current levels of 100 mA
1-3. Control of the 5205 A can be exercised either manually
from the front panel controls or by remote programming.
Remote connectors for the 5200A and a customer supplied
control source are provided on the rear panel. Connection of
the Model 5200A AC Calibrator with the supplied factory
cable provides all the analog signal and remote programming
data required.
1-4. The amplifier output cable and the input for
independent operations are installed on the front panel.
Both may be installed on the rear panel for system
operations if desired as option —07. The standard output
cable is a permanently attached, three foot long, four wire,
low capacitance cable which allows remote sensing, high
output current and high frequency operation.
1-5. Equipment protection is provided by the Fault and the
Overload circuits. Activating the fault circuit
automatically reverts the unit to the Standby mode and
removes any voltage present from the output. Faults can
originate from procedureal errors or equipment failure. A
fault due to procedural error can be removed by resetting
the systems and performing the restarting procedure. An
overload error can be corrected by removing the overload
from the circuit. Return to operate will occur automatically
when the overload is removed, unless the overload is
accompanied by a fault. In this case, power must be removed
and restored to remove the indications.

These manuals are available for the above equipment:

Fluke -- 5205A -- Service and User Manual
File name: Fluke-7679-Manual-Page-1-Picture
Manual Type: Service and User Manual
Pages: 132
Size: 5.49 Mbytes (5755224 Bytes)
Language: english
Manual-ID/Number: 338434
Date: April 1976
Quality: Scanned document, all readable.
Upload date:
MD5: 288ea59ced53095c96169c40c130e7a8
ed2kID: 6ac3ea6bdc84d7c39a79967648a10fc3
Downloads: 34 since 06 February 2016
No Schematics!
Start Free Download


No comments yet.

Add Comment

* Required information
(never displayed)
Bold Italic Underline Strike Superscript Subscript Quote Line Bullet Link Email Image Video
Smile Sad Huh Laugh Mad Tongue Crying Grin Wink Scared Cool Sleep Blush Unsure Shocked
Enter the word table backwards.
Enter answer:
Notify me of new comments via email.
Remember my form inputs on this computer.
Powered by Commentics