Equipment information:

Fluke 4265A Power supply
Manufacturer:
Fluke
Model: 4265A
Date: 1976
Category: Measurement instruments
Group: Power supply
Description: Binary Programmable Power Source
Information:
                                    The Model 4265A is a programmable, bipolar dc voltage
source. Output voltage is from 0 to 65.532 vdc in two
ranges; 16 and 65 vdc. Programming resolution using the
internal reference voltage is 1 mv on the 16 volt range and
4 mv on the 65 volt range. Output accuracy is ±.01% of the
programmed level on both voltage ranges. After a 100 /sec
settling time, a READY/NOT READY FLAG indicates that the
output has settled to ±.01% of the programmed increment for
a resistive load. The output current is rated at ±1 ampere
maximum and is limited to approximately ±1.1 ampere in the
event of an overload or short circuit. The sink current
(milliamps) is rated at 1000-10 Eout maximum. A Programmable
Current Limiter (Option -06) is available to allow
programming the maximum output current to a lower level. A
current limit flag indicates when a sink or source current
overload exists with or without the -06 Option.
1-3. Programming requirements are compatible with DTL or TTL
logic levels. Contact or relay closures can also be used.
The voltage source is programmed on the low range using
straight binary coding. The high range is also programmed in
the same manner, but the actual output is four times the
programmed level. Two's complement coding is required for
negative outputs when the Isolated Control Logic (—01
Option) is installed. All programming inputs
and flag outputs are available through a 50 pin Amphenol,
Blue Ribbon connector located on the rear panel. A +5 vdc
output is also available at this connector and is used to
provide power for external programming circuitry. Negative
logic is employed for programming. The logic levelsi are as
follows:
Logic “0” = +2.0 vdc to +5.0 vdc or open circuit
Logic “1” = 0 to +0.4 vdc or short circuit to LOGIC GRD.
1-4, Three options are available to provide tailoring of the
power supply to fit application requirements. These options
are identified by numeric designations -03, -06, and -09. A
description of each option is provided in Table 1-1. These
options may be installed at the factory when the instrument
is ordered or in the field at a later time.
1-5. Physically, the power source is completely solid state.
Plug in printed circuit boards, with easy to reach test
points and adjustments, are used for ease in servicing. The
unit is forced air cooled. This results in lower component
temperatures and thus higher reliability than would normally
be obtained in a source of this power capability. The
chassis is designed for bench top use, or it can be
installed in a standard equipment rack by using the
Accessory Rack Mounting Fixtures.

These manuals are available for the above equipment:

Fluke -- 4265A -- Service and User Manual
File name: Fluke-7673-Manual-Page-1-Picture
Manual Type: Service and User Manual
Pages: 120
Size: 4.65 Mbytes (4873699 Bytes)
Language: english
Revision: 1
Manual-ID/Number: 310631
Date: February 1976
Quality: Scanned document, all readable.
Upload date:
MD5: cc08fc49321b70b3983c15fb5ce9bda2
ed2kID: 736bd82c81230b48fd334d5ff0326b02
Downloads: 38 since 31 January 2016
Information:
1 INTRODUCTION AND
SPECIFICATIONS.............................................1-1
1-1. Introduction.......................
..............................................1-1
1-6.
Specifications.............................................................................1-2
2 OPERATING
INSTRUCTIONS...............................................................2-1
2-1.
Introduction.............................................................................2-1
2-3. Shipping
Information...............................................................2-1
2-6. Input
Power.............................................................................2-1
2-8. Rack
Installation........................................................................2-1
2-10. Operating
Features....................................................................2-3
2-12. Operating
Notes........................................................................2-3
2-24, Programming
Information...........................................................2-4
3 THEORY OF
OPERATION....................................................................3-1
3-1.
Introduction.............................................................................3-1
3-3. Fundamental Circuit
Description.............................................3-1
3-5. Block Diagram
Analysis...........................................................3-2
3-14. Circuit Descriptions
....................................................................3-3
4
MAINTENANCE......................................................................................4-1
4-1.
Introduction..............................................................................4-1
4-3. Service
Information....................................................................4-1
4-6. General
Maintenance...............................................................4-1
4-13. Maintenance
Access....................................................................4-1
4-15. Calibration
Procedures...............................................................4-2
5 LIST OF REPLACEABLE
PARTS.................................................
5-1
Introduction...................................................................
54 Parts List Column
Descriptions........................................
5-5 Manufacturers’ Cross Reference List Column Descriptions .
5-6 How to Obtain
Parts......................................................
5-9 Serial Number
Effectivity..................................................
6 OPTIONS AND ACCESSORIES
..........................................................
6-1
Introduction....................................................................
6-3 Option
Information..........................................................
6-5 Accessory
Information......................................................
Model A4200 Manual Control Unit Schematic Diagram A4200-1000
Option -09, Multi-Strobe Isolated Logic Schematic Diagram
4200-1024 (2 sheets)
Schematic Diagram 4200-1025 Rack Mounting Fixtures Program
Connector
Extender PCB assembly and Cable Extender Assembly
7 GENERAL
INFORMATION...........................................................
8 SCHEMATIC
DIAGRAMS...........................................................
A1 Mother Board/A7 Binary Display
4265 A-1011
4265A-1013
A2 Power Supply/A10 Connector Board
4250A-1061
4250A-1012
A3 Preamplifier
4265A-1051
A4 Option -03, External Reference
4210A-1041
A5 Binary Ladder
4216A-1031
A6 Isolated Control Logic (2 sheets)
4275A-1021
A8 Option -06, Programmable Current Limiter
4250A-1081
A9 Power Amplifier
4250A-1071
Fluke-7673-Manual-Page-1-Picture
Start Free Download


Comments

No comments yet.

Add Comment

* Required information
(never displayed)
 
Bold Italic Underline Strike Superscript Subscript Quote Line Bullet Link Email Image Video
 
Smile Sad Huh Laugh Mad Tongue Crying Grin Wink Scared Cool Sleep Blush Unsure Shocked
 
1000
Enter the third word of this sentence.
 
Enter answer:
 
Notify me of new comments via email.
 
Remember my form inputs on this computer.
 
 
Powered by Commentics