CATHODE-RAY OSCILLOGRAPH
Information
INTRODUCTION
The Du Mont Type 304-A is a general-purpose cathode-ray
oscillograph that enables rapid, accurate amplitude
measurements of any portion of a 0-to 1000-volt signal, from
d-c to over 50,000 cps.. The high gain assures that the Type
304-A can be used directly with most transducers, while d-c
amplification provides that the lowest frequency portions of
signals will be faithfully reproduced.
To obtain the utmost from the Type 304-A, full advantage
should be taken of the novel amplitude calibrating system of
the instrument which permits signal amplitude measurements
in volts directly from the screen. (See "Amplitude
Calibration.") Unlike electro-mechanical devices, the Type
304-A is not restricted to rms or peak-to-peak readings of
voltage, as any amplitude portion of the input signal may be
measured with the instrument. Moreover, these measurements
may be made in a minimum of time, qualifying the Type 304-A
as a quantitative as well as a qualitative instrument.
The TYPE 5ADP-Cathode-ray Tube is employed in the Type
304-A. This tube is precision-built to tolerances far more
stringent than in conventional tubes, and provides a
deflection sensitivity up to twice that of conventional
types, as well as a smaller spot size with no sacrifice in
brilliance. The Types 304-A and 304-AR differ only in that
the -Type 304-AR is designed for mounting in a standard,
19-inch relay rack.
2. PHYSICAL CHARACTERISTICS
Physical specifications of both the Type 304-A and the Type
304-AR are given in convenient tabular form in Table 1-1.
The Type 304-A is enclosed in a blue-gray metal cabinet
complete with a convenient carrying handle. The Type 304-AR
employs a standard relay rack panel, a dust cover and a
bottom plate. The operating controls of both instruments are
conveniently located on the front panel with vertical
Amplifier controls on the left and horizontal sweep controls
on the right. In both instruments a convenient terminal
panel, accessible at the rear, provides direct access to the
deflection plates of the cathode-ray tube.
To facilitate calibration, the scale of the Type 304-A is
numbered and edge illuminated, with a dimmer control
provided on the front panel. Unidirectional signal
amplitudes are measured with the aid of the left-hand scale,
symmetrical signals with the aid of scale on the right hand.
These scales make it especially convenient to photograph a
trace with scale superimposed. The usual double-exposure
technique is normally not necessary when employing an
instrument with an illuminated scale; thus, valuable time is
saved, and more accurate results are obtained. The Type
304-A is also provided with a suitable color Filter over the
CRT screen to improve visual contrast.
Provision is made for connection of balanced, as well as
single-ended vertical input signals on the 0.1 VOLT FULL
SCALE range. Balanced input may be obtained merely by
disconnecting a jumper to ground from the lower Y-input
terminal on the front panel.
Dokumenttyp:
Service- und Bedienungsanleitung
Seiten:
64
Größe:
7.33 Mbytes (7680985 Bytes)
Sprache:
english
Revision:
Dokument-ID:
Datum:
Qualität:
Gescanntes Dokument, teilweise schwer oder teilweise nicht zu lesen.
Upload Datum:
2017 05 07
MD5:
91faff993bc41a24550302086697aba0
Downloads:
1109
Information
1 INTRODUCTION ... 1
2 PHYSICAL CHARACTERISTICS ... 1
3 CIRCUITRY EMPLOYED ... 3
SECTION II OPERATION
1 UNPACKING AND INSPECTING THE EQUIPMENT ... 10
2 LOCATION AND INSTALLATION ... 10
a. General ... 10
b. Line Voltage and Frequency ... 10
c. Power Regulation ... 10
3 TURNING ON THE EQUIPMENT ... 11
4 PRECAUTION AGAINST SCREEN BURNING ... 11
5 CONTROLS AND TERMINALS ... 11
a. General ... 11
b. Control and Terminal Functions . . ... 11
c. Synchronization of Sweep ... 12
d. Obtaining Low-frequency Sweeps . ... 15
e. Amplitude Calibration ... 16
f. Intensity Modulation ... 17
g. Direct Connection To Deflection Plates ... 18
6 PHOTOGRAPHING OSCILLOGRAMS ... 19
7 USE OF BALANCED INPUT ... . 20
SECTION III THEORY OF OPERATION
1 GENERAL ... 22
2 CATHODE-RAY TUBE CIRCUIT ... 22
a. General ... 22
b. Intensity ... 23
c. Focus ... . . ... 23
d. Intensity Modulation ... 23
3 VERTICAL DEFLECTION CIRCUIT (Y AXIS) ... 23
a. General ... 23
b. Input Attenuator ... 23
c. Voltage Calibrator ... 24
d. Vertical (Y-axis) Amplifier ... 24
4 SYNC AND SWEEP CIRCUITS ... 26
a. General ... 26
b. Synchronization ... 26
c. Recurrent Sweep ... 28
d. Driven Sweep ... 28
5 HORIZONTAL DEFLECTION CIRCUIT (X AXIS) ... 29
a. General ... 29
b. Circuit Theory ... 29
6 POWER SUPPLY ... 31
a. General ... 31
b. Functions ... 31
c. Low-Voltage Power supply ... 31
d. High-Voltage Power supply ... 31
e. Regulated Heater Supply ... ... 33
2 THEORY OF LOCALIZATION ... 33
3 CIRCUIT VOLTAGES AND RESISTANCES ... 38
4 SELECTING 115- or 230- VOLT LINE ... 40
5 FACTORY ADJUSTMENTS ... 41
a. General ... 41
b. Test Equipment Required. . ... 41
c. Factory Adjustments for Y Amplifier ... 41
d. Factory Adjustments for X Amplifier ... 48
e. Factory Adjustments for the Sweeps ... 49
g. Other Factory Adjustments ... 50
6 REPLACEMENT OF CATHODE-RAY TUBE . ... 51
a. Type 304-A ... 51
b. Type 304-AR ... 51
7 ILLUMINATED CALIBRATED SCALE ... 51
8 DU MONT "OSCILLOGRAPHER" ... 52
9 DU MONT WARRANTY ... 52
10 SERVICE ... 52