Geräte Informationen:

Racal 2151 Zähler
Hersteller:
Racal
Modell: 2151
Datum: 1991
Kategorie: Messgeräte
Gruppe: Zähler
Beschreibung: VXI Counter
Informationen:
                                    1 The Racal Instruments VXlbus microwave counter, Model
2151, is a microprocessor- controlled instrument offering
high-accuracy measurements with a comprehensive range of
facilities. These facilities include Automatic/Manual
Operation, Tracking, Offset, Multiply, and Store and Recall.
All the instrument’s functions are controlled remotely via
the VXI interface.
2 The Racal Instruments VXlbus UHF counter, Model 2051,
offers high accuracy measurement over a reduced frequency
range, with a similar range of facilities to those provided
on the Model 2151.
MEASUREMENT FUNCTIONS
Frequency A Function
3 The Frequency A Function is used to measure the frequency
of the signal applied to INPUT A. A resolution of up to ten
digits can be selected with a gate time of 1 millisecond to
20 seconds depending on the resolution selected.
Frequency B Function
4 The Frequency B Function is used to measure the frequency
of the signal applied to INPUT B. A resolution of up to ten
digits can be selected with a gate time of 1 millisecond to
20 seconds depending on the resolution selected.
Frequency C Function (2151 only)
5 The Frequency C Function is used to measure the frequency
of the signal applied to INPUT C. A resolution of 0.1 Hz to
1 MHz can be selected with a gate time of 1 millisecond to
10 seconds depending on the resolution selected and
frequency being sampled.
Ratio C/B Function (2151 only)
6 The Ratio C/B Function is used to measure the ratio of the
frequency applied to INPUT C to that applied to INPUT B.
Ratio C/A Function (2151 only)
7 The Ratio C/A Function is used to measure the ratio of the
frequency applied to INPUT C to that applied to INPUT A.
Ratio B/A Function
8 The Ratio B/A Function is used to measure the ratio of the
frequency applied to INPUT B to that applied to INPUT A.

Folgende Anleitungen sind für obiges Gerät verfügbar:

Racal -- 2151 -- Serviceanleitung
Dateiname: Racal-7287-Manual-Page-1-Picture
Dokumenttyp: Serviceanleitung
Seitenanzahl: 66
Dateigröße: 578.33 Kbytes (592210 Bytes)
Sprache: Englisch
Revision:
Dokument-ID/Nummer:
Datum: März 1991
Qualität: Gescanntes Dokument, alles ist lesbar.
Upload Datum:
MD5: be2c3aa1dad39bae6c7982e79f615f42
ed2kID: 307ea4d04c018f79e321b1c15610e7d2
Downloads: 21 seit 10 Oktober 2015
Informationen:
TECHNICAL
SPECIFICATION............................................................................1-1
GENERAL DESCRIPTION
............................................................................2-1
INTRODUCTION.........................................................................................................2-1
MEASUREMENT FUNCTIONS
............................................................................2-1
Frequency A Function
......................................................................................2-1
Frequency B Function
......................................................................................2-1
Frequency C Function (2151
only)..................................................................2-1
Ratio C/B Function (2151 only)
..................................................................2-1
Ratio C/A Function (2151 only)
..................................................................2-1
Ratio B/A
Function...............................................................................................2-1
CHECK FUNCTION
...............................................................................................2-2
SIGNAL INPUT CHANNELS
............................................................................2-2
ADDITIONAL FREQ C ACQUISITION MODES (2151 ONLY)
..................2-3
MATHS FUNCTIONS
...............................................................................................2-3
SPECIAL
FUNCTIONS...............................................................................................2-3
STORE/RECALL FUNCTIONS
...........................................................................2-3
SAMPLE RATE
.........................................................................................................2-3
ERROR INDICATION
...............................................................................................2-4
HOLD FEATURE............
......................................................................................2-4
RESOLUTION AND GATE
TIME...........................................................................2-4
EXTERNAL FREQUENCY STANDARD
INPUT..............................................2-4
INITIALIZATION........................................................................................................2-4
VXI
INTERFACE........................................................................................................2-5
PREPARATION FOR
USE.....................................................................................3-1
UNPACKING
........................................................................................................3-1
SETTING THE LOGICAL ADDRESS AND INTERRUPT
PRIORITY
SWITCHES..............................................................................................3-1
INSTALLATION
........................................................................................................3-1
Connection to the VXI
.....................................................................................3-1
FREQUENCY
STANDARD.....................................................................................3-1
FUNCTIONAL
CHECK............................................................................................3-5
VXI Check
........................................................................................................3-5
CHE CK Message
Check.....................................................................................3-5
Data Output Check
*.....................................................................................3-6
SRQ and Status Byte Check
...........................................................................3-6
Input Channel Function Checks
.................................................................3-7
OPERATION VIA THE VXI
...........................................................................4*1
INTRODUCTION
..............................................................................................4-1
DESCRIPTION OF FRONT PANEL CONNECTORS
AND INDICATORS ..............................................4-1
DESCRIPTION OF SIDE PANEL SETTING CONTROLS
...........................4-2
INITIAL SWITCH-ON
..............................................................................................4-3
FREQUENCY MEASUREMENT- INPUT
A.......................................................4-3
FREQUENCY MEASUREMENT - INPUT
B.......................................................4-4
FREQUENCY MEASUREMENT - INPUT C (2151 ONLY)
...........................4-4
Normal Automatic Mode
..........................................................................4-4
Manual
Mode.......................................................................................................4-5
Further Functions in FREQ C
(Automatic Mode
Only)....................................................................................4-5
Set LOW FM (FREQ
C)....................................................................................4-5
Set TRACK (FREQ C)
...................................................................................4-6
RATIO MEASUREMENTS C/B (2151 ONLY),
C/A (2151 ONLY) AND B/A
.........................................................................4-6
MEASUREMENT SAMPLE RATE
................................................................4-7
Setting and U sing the Sample
Rate................................................................4-7
Interrogating the Sample Rate
................................................................4-7
MULT
FUNCTION......................................................................................................4-7
Setting the Multiplier
...................................................................................4-7
Enabling the
Multiplier...................................................................................4-7
Interrogating the
Multiplier..........................................................................4-7
OFFSET FUNCTION
.............................................................................................4-8
Setting the
Offset.............................................................................................4-8
Enabling the Offset
...................................................................................4-8
Interrogating the
Offset...................................................................................4-8
SMOOTH FUNCTION......................
................................................................4-8
CLK10 FREQUENCY STANDARD
................................................................4-8
GATE
TIME.................................................................................................................4-9
SPECIAL FUNCTIONS
...................................................................................4-10
Setting the Special Function Register
.......................................................4-11
Enabling and Disabling the Special Functions
...................................4-11
ERROR CODES ............
...................................................................................4-12
Clearing the Error Codes
.........................................................................4-12
VXI OPERATING MODES
...................................................................................4-13
READING OUT
DATA.............................................................................................4-13
CONTROLLING THE
INSTRUMENT................................................................4-13
DATA OUTPUT
FORMAT-NORMAL........................................
................4-14
DATA OUTPUT FORMAT -.DUMP
MODES.......................................................4-14
REQUESTING SERVICE
...................................................................................4-18
STATUS BYTE
.......................................................................................................4-18
SETTING THE ENABLE
REGISTERS,................................................................4-22
MESSAGE PROTOCOLAND SYNTAX
......................................................4-22
System Message
Traffic...................................................................................4-22
Program
Messages.............................................................................................4-22
Response
Messages............................................................................................4-23
White Space
Character..................................................................................4-23
OVERLAPPED/SEQUENTIAL COMMANDS
.............................................4-23
COMMANDS
......................................................................................................4-24
Racal-7287-Manual-Page-1-Picture
Kostenlos herunterladen


Kommentare

Noch keine Kommentare vorhanden.

Kommentar Hinzufügen

* Pflichtfeld
(wird nicht veröffentlicht)
 
Bold Italic Underline Strike Superscript Subscript Quote Line Bullet Link Email Image Video
 
Smile Sad Huh Laugh Mad Tongue Crying Grin Wink Scared Cool Sleep Blush Unsure Shocked
 
1000
Enter the word table backwards.
 
Die Antwort lautet:
 
Informiere mich via E-Mail über neue Kommentare.
 
Meine Daten auf diesem Computer.
 
 
Powered by Commentics